Product Info

Technical data: - Half value layer (HVL): 2,2 mm Al. - Total filtration at 70 kV: 2,4 mm Al. - Tube inherent filtration at 70 kV: > 1 mm Al. - Focal spot size: 0,7 mm. - X-ray tube tension accuracy: ±10%. - X-ray tube current accuracy: ±20%. - Radiation linearity: ±10%. - X-ray emission time accuracy: ±20 ms 0,020 s ≤ t ≤ 0,320 s, ±5% 0,400 s ≤ t ≤ 3,2 s. - Reproducibility: 0,05. - Generator: At constant potential. - X-ray tube nominal current: 4 mA / 8 mA. - X-ray tube nominal voltage: 60 / 70 kV. - Exposure times: 0,020 s ÷ 3,2 s (23 steps). - Reference current-time product: 0,8 mAs 8 mA 0,1 s, 0,4 mAs 4 mA 0,1 s. - Intensity of radiation in the air: > 30 μGy/h at 1 m from focal spot. - Leakage radiation (measured @ 70kV, 8 mA, 3,2 s): < 0,25 mGy/h at 1 m from focal spot - Operating cycle: 1:32. - Focal spot: 0.7 mm.

X-Ray Wall Mounted DC 0.7 mm

1 Unit

Organise to Lists

Place this product into lists for easy-ordering

Add to List